Technical Specifications

Model Dimensions Display
N50-GO-spectrophotometer-implen-Go-nanophotometer-nanodrop-alternative N50-GO-dimensions-implen-GO-nanophotometer-spectrophotometer-nanodrop-alternative Implen-GO-N50-GO touchscreen display
N50-GO 200 mm x 200 mm x 120 mm 1024 x 600 pixels
Glove compatible touchscreen
Processing Power Weight Hardware & Software
2.4 GHz
Intel Celeron dual core processor
4.5 kg • Linux based NPOS
• 32 GB Internal Data Storage
• 2x USB A, USB B, HDMI, Ethernet, WiFi
• Windows 8, 10 (32 & 64 bit)
OS X (Intel x86 and Apple M1)
iOS and Android OS
NanoVolume Performance Detection Range dsDNA 5 – 7,500 ng/µl
Detection Range BSA 0.15 – 217 mg/ml
Minimum Sample Size 0.3 – 2 µl
Photometric Range (10 mm equivalent) 0.1 – 150 A
Path Length 0.67 and 0.07 mm
Dilution Factor 15 and 140
Optical Specifications Wavelength Scan Range 200 – 650 nm
Measure Time For Full Scan Range 2.5 – 4.0 seconds
Wavelength Reproducibility ± 1 nm
Wavelength Accuracy 1.5 nm
Bandwidth < 3 nm
Stray Light < 0.5% at 220 nm using NaI
Absorbance Reproducibility (Lid 15): < 0.004 A @ 0 – 0.3 A @ 280 nm
CV < 1% @ 0.3 – 1.5 A @ 280 nm
Absorbance Accuracy < 1.75% @ 0.7 A @ 280 nm of the reading
Stray Light < 2% @ 240 nm using NaI
Optical Arrangement 1 x 4096 CMOS Array
Lamp Xenon flash lamp
Lifetime 109 flashes, up to 10 years
General Specifications Operating Voltage 90 – 250 V, 50/60 Hz, 90 W, 18/19 VDC
Certification CE, IEC 61010-1:2012 and EN 61326-1:2013
Security Slot for Kensington lock

Features and specifications are subject to change without notice.